A2Z Ozone Bed Sterilizer User Manual
Page 11

11
How do you handle problems during use of Bed Sterilizer?
Malfunction
Reason
Suggestions
No electricity after main
switch is on
No power supply enter
into device
1. Check if the power plug to
be used is the correct voltage
for the unit.
2. Check if the socket has
electricity.
3. Check the fuse.
1. Could not be startup
the unit after power is on
2.Alarm
rings
continuously
3. Abnormal display
Chipset malfunction
1. Turn off the main switch,
and restart it after 1 minute.
Contact manufacture if
it’s not
working.
2. Do not try to repair it
without professional
assistance.
Very strong ozone smell
Ozone leakage
1. Check the hose, nozzle and
sealed bags. If there is any
leakage problem, please
change it.
2. If any inner parts have a
leakage problem, contact us.
No ozone generating
when in operation
Ozone cell
malfunction.
Contact us.