beautypg.com

3 electrical requirements, 4 installing a platform, Electrical requirements – Eppendorf Excella E-10 Classic Platform Shaker User Manual

Page 12: Installing a platform

background image

Installation
New Brunswick Excella® E-10 Open Air Shaker
English (EN)

12

Operating Environment

The shaker is designed to operate optimally in the following ambient conditions:

• 5 – 45 °C
• 20 – 80 % Relative Humidity non-condensing
• Up to 2000 m

4.3

Electrical requirements

The Excella E-10 can be equipped to run on:

• 100/120 volts, 50/60 Hz, 1500 VA maximum
• 230 volts, 50/60 Hz, 1500 VA maximum

In all cases, voltage variations must not exceed ±10 %.

4.4

Installing a platform

The shaker is shipped with the four Allen (hexagonal, countersunk) head platform screws installed in the
upper bearing housing. These screws must be removed before installing a platform. The Allen key is
provided.

A platform must be installed on the shaker prior to use.

WARNING!

Unplug the shaker before installing the platform.