5 additional data, 1 interfaces, 2 time interval – Eppendorf Thermomixer R User Manual
Page 28: 1 interfaces 9.5.2 time interval, 9 technical data
9 Technical data
30
Thermomixer comfort / Thermomixer R — Operating Manual
9.5 Additional data
9.5.1
Interfaces
9.5.2
Time interval
Heating rate:
approx. 5 °C/min
The temperature changes in the filled tubes
are slower.
Cooling rate:
2 to 3 °C/min
Cooling rate from room temperature to 13 °C
below room temp:
0.5 to 1.0 °C/min
Mixing frequencies for:
Exchangeable thermoblocks 0.5 ml
300 to 1,500 1/min
Exchangeable thermoblocks 1.5 and 2.0 ml
300 to 1,400 1/min
Microplates
300 to 1,400 1/min
Falcon tubes
300 to 750 1/min
Cryo tubes
300 to 1,400 1/min
EDP connection:
RS-232, Sub-D9 male
Hint!
Only connect devices to interfaces that meet the IEC 950/EN 60950 (UL 1950)
standards.
Programmable time interval:
1 min to 99:59 hours, infinitely adjustable
EN