Technical data, 1 power supply, 2 ambient conditions – Eppendorf Thermomixer R User Manual
Page 27: 3 weight / dimensions, 4 application parameters, 9 technical data
9 Technical data
29
Thermomixer comfort / Thermomixer R — Operating Manual
9
Technical data
9.1 Power supply
9.2 Ambient conditions
9.3 Weight / dimensions
9.4 Application parameters
Mains connection:
100 to 240 V ±10 %, 50 to 60 Hz
Fuses:
see nameplate or label on the rear of device
Power consumption:
90W
Environment:
For indoor use only
Ambient temperature:
4 to 35 °C
Relative humidity:
70 % max.
Dimensions (W x D x H):
220 x 250 x 125 mm
Weight basic device:
3.2 kg
Temperature control range:
Exchangeable thermoblocks for micro test tubes 13 °C below room temperature to 99 °C
Exchangeable thermoblocks for micro test
plates
10 °C below room temperature to 99 °C
Temperature accuracy:
Exchangeable thermoblock for 1.5 ml micro test
tubes
at nominal values between 20 °C and 45 °C
±0.5 °C
at nominal values between <20 °C and >45 °C ±2.0 °C
Exchangeable thermoblocks for micro test
plates
at nominal values <70 °C
±2.0 °C
at nominal values between 70 °C and 99 °C
-5.0 °C
EN
Operating Manual