9 Technical data

29

Thermomixer comfort / Thermomixer R — Operating Manual

9

Technical data

9.1 Power supply

9.2 Ambient conditions

9.3 Weight / dimensions

9.4 Application parameters

Mains connection:

100 to 240 V ±10 %, 50 to 60 Hz

Fuses:

see nameplate or label on the rear of device

Power consumption:

90W

Environment:

For indoor use only

Ambient temperature:

4 to 35 °C

Relative humidity:

70 % max.

Dimensions (W x D x H):

220 x 250 x 125 mm

Weight basic device:

3.2 kg

Temperature control range:

Exchangeable thermoblocks for micro test tubes 13 °C below room temperature to 99 °C

Exchangeable thermoblocks for micro test

plates

10 °C below room temperature to 99 °C

Temperature accuracy:

Exchangeable thermoblock for 1.5 ml micro test

tubes

at nominal values between 20 °C and 45 °C

±0.5 °C

at nominal values between <20 °C and >45 °C ±2.0 °C

Exchangeable thermoblocks for micro test

plates

at nominal values <70 °C

±2.0 °C

at nominal values between 70 °C and 99 °C

-5.0 °C