IKA C 6000 isoperibol Package 2/12 User Manual
Page 28

28
Click on the necessary fi gures and the decimal
point, and press OK to confi rm the input.
You can also load the weight from a connected
balance or request it by clicking on the „Balance“
button. The value is then entered automatically.
All other inputs are optional.
The default is that the decomposition vessel is not
selected manually but is identifi ed when it has
been hung in the calorimeter.
Note: weight must be < 5 g.
An alphanumeric virtual keyboard appears for
the input options „Designation“, „User“, „Proper-
ties“:
Click on OK to complete the respective input.
There is a special procedure for the designation:
the default is that the designation for a sample is
constructed from the current date and a sequen-
tial number. For instance JJMMTT07 denotes the
seventh sample processed on TT.MM.JJ.
This designation is also displayed the next time. If
the last characters of the designation is numeric,
this is incremented by one.
You can stop the use of the user-specifi c desig-
nation again every time: call up the „Designati-
on“ dialogue, delete the designation and click on
„OK“.
Highlight the „Calibration“ fi eld if you wish to
perform a calibration.
The scroll bar allows you to access additional in-
put fi elds for inputting the external energies. If no
special application procedures must be observed
you can here specify use of the „IKA
®
ignition
thread“, „User-defi ned“ = 0, „User-defi ned“ = 0
default settings. This means that the default IKA
®
ignition thread will be used as the ignition aid and
that no further combustion aids will be used.
Simulation
In some circumstances it may be helpful to check
some calorifi c value tests or to calculate possi-
ble test results without actually performing the
combustion test. This is particularly useful when
a calibration has been carried out instead of a ca-
lorifi c value determination or vice versa. This can
be corrected by simulation using the increase in
temperature from the wrongly interpreted mea-
surement. The decomposition vessel used can be
selected after activating the simulation.
Working with the IKA
®
C 6000 global standards/isoperibol calorimeter