Engineering
Standards

Name

Identifier

Instructions

Engineering Standard Number

C1-102190-A

Digital Pressure Gauge

Digital Display Calibration should be performed to assure accuracy.

Display Calibration sequence for Gauges with 7 segment LED’s

A password will look like the following example.

L

EFT

R

IGHT

R

IGHT

L

EFT

L

EFT

Enter the switch sequence with a magnet to enter the basic calibration mode.

If the password is correctly entered,

CAL

will be displayed for 1/2 second followed by a num-

ber that is dependent on the display being calibrated (usually

0

). I

Establish accurate and stable conditions for calibration.

When the calibration conditions are stable, activate the left switch.

The display will read

donE

and then return to normal operation.