Programming overcurrent protection, 1 turn on the power supply, 2 enable the output – Agilent Technologies E3632A User Manual
Page 47: 3 enter the ocp menu and set the trip level, 4 enable the ocp circuit, Programming overcurrent protection 45, Setting the ocp level and enable the ocp circuit

Chapter 3 Front-Panel Operation
Programming Overcurrent Protection
45
3
Programming Overcurrent Protection
Overcurrent protection guards the load against output currents that reach a
specified value greater than the programmed protection level. It is
accomplished by programming the output current to zero.
The following steps show how to set the overcurrent protection trip level, how
to check OCP operation and how to clear overcurrent condition.
• Front-panel operation:
Setting the OCP Level and Enable the OCP Circuit
1
Turn on the power supply.
The power supply will go into the
power-on / reset
state; the output is disabled
(the
OFF
annunciator turns on); the 15V/7A range is selected (the
15V
annunciator turns on); and the knob is selected for
voltage
control.
2
Enable the output.
The
OFF
annunciator turns off and the display will go to the meter mode.
3
Enter the OCP menu and set the trip level.
You will see the above message on the display when you enter the OCP menu.
Adjust the knob for the desired OCP trip level.
4
Enable the OCP circuit.
You will see the above message after pressing the
key.
LEVEL 7.5 A
OCP ON
Power
Output On/Off
Over Current
Over Current
Over Current