Performance specifications, Performance specifications 157 – Agilent Technologies E3632A User Manual
Page 160

Chapter 8 Specifications
Performance Specifications
157
8
Performance Specifications
Output Ratings(@0°C - 40°C)
Low range
0 to +15 V/0 to 7 A
High range
0 to +30 V/0 to 4 A
Programming Accuracy
[1]
12 months(@25°C ± 5°C), ±(% of output + offset)
Voltage
0.05% + 10 mV
Current
0.2% + 10 mA
Readback Accuracy
[1]
12 months(over GPIB and RS-232 or front panel with respect
to actual output @25°C ± 5°C), ±(% of output + offset)
Voltage
0.05% + 5 mV
Current
0.15% + 5 mA
Ripple and Noise (with outputs ungrounded, or with either output terminal grounded,
20 Hz to 20 MHz)
Normal mode voltage
<0.35 mV rms and 2 mV p-p
Normal mode current
<2 mA rms
Common mode current
<1.5
μA rms
Load Regulation, ±(% of output + offset)
Change in output voltage or current for any load change within ratings with remote
sensing connected
Voltage
<
0.01% + 2 mV
Current
<
0.01% + 250
μA
Line Regulation, ±(% of output + offset)
Change in output voltage and current for any line change within ratings
Voltage
<
0.01% + 2 mV
Current
<
0.01% + 250
μA
[1]
Accuracy specifications are after an 1-hour warm-up with no load and calibration at
25 °C.