6 return to the meter mode, 7 enable the output, Note – Agilent Technologies E3632A User Manual
Page 41: Current, Blinking, Meter, Annunciators, Higher

Chapter 3 Front-Panel Operation
Constant Current Operation
39
3
5
Adjust the knob for the desired output current.
1
Check that the
Limit
annunciator still blinks. Set the knob for
current
control.
The second digit of the ammeter will be
blinking
. Change the blinking digit
using the resolution selection keys and adjust the knob to the desired output
current.
6
Return to the meter mode.
Press
key or let the display time-out after several seconds to
return the meter mode. Notice that the
Limit
annunciator turns off and the
display shows “OUTPUT OFF” message.
7
Enable the output.
The
OFF
annunciator turns off and the
15V
or
30V
,
OVP
,
OCP
and
CC
annunciators are lit. Notice that the display is in the meter mode. In the
meter
mode, the display shows the actual output voltage and current.
Refer to “Programming Overvoltage Protection” and “Programming
Overcurrent Protection” sections, starting on page 42 and page 45 for more
information on
OVP
and
OCP
annunciators.
8
Verify that the power supply is in the constant current mode.
If you operate the power supply in the constant current (CC) mode, verify that
the
CC
annunciator is lit. If the
CV
annunciator is lit, choose a
higher
voltage
limit.
N o t e
During actual CC operation, if a load change causes the voltage limit to be
exceeded, the power supply will automatically crossover to constant voltage
mode at the preset voltage limit and the output current will drop
proportionately.
• Remote interface operation:
VOLTage {
Set the voltage
CURRent {
Set the current
OUTPut ON
Enable the output
1
You can use the resolution selection keys to move the blinking digit to the
right or left when setting the current.
Volt/Curr
Display Limit
Display Limit
Output On/Off