Command separators 113, Using the, Command separators – Agilent Technologies E3632A User Manual
Page 116: Parameters

Chapter 4 Remote Interface Reference
An Introduction to the SCPI Language
113
4
Command Separators
A
colon
( : ) is used to separate a command keyword from a lower-level
keyword as shown below:
"SOURce:CURRent:TRIGgered"
A
semicolon
( ; ) is used to separate two commands within the same subsystem,
and can also minimize typing. For example, sending the following command
string:
"SOUR:VOLT MIN;CURR MAX"
... is the same as sending the following two commands:
"SOUR:VOLT MIN"
"SOUR:CURR MAX"
Use a colon
and
a semicolon to link commands from
different
subsystems.
For example, in the following command string, an error is generated if you do
not use the colon and semicolon:
‘‘
DISP:TEXT:CLE;:SOUR:CURR MIN
’’
Using the
MIN
and
MAX
Parameters
You can substitute
MINimum
or
MAXimum
in place of a parameter for many
commands. For example, consider the following command:
CURRent {<current>|MIN|MAX}
Instead of selecting a specific current, you can substitute
MINimum
to set the
current to its minimum value or
MAXimum
to set the current to its maximum
value.