Instruction manual ni-205 – WIKA APA10 User Manual
Page 4

INSTRUCTION MANUAL
NI-205
Rev. 1 12/02
 
The pressure switch must be mounted in the normal 
installation position, i.e. with the pressure connection 
pointing downwards. 
Fig. 6 -
Calibration circuit
Pressione 
assoluta 
PS - Pressure switch 
CA - Test pressure gauge for 
absolute pressuures
V1 - Inlet valve
V2 - Outlet valve 
P - 
Absolute
pressure
source
Avoid forcing the elastic support of the microswitch by 
hand or with tools. This could affect the instrument 
functioning. 
CAUTION: if the switch is of the kind with adjustable 
dead band (letter R of contacts code) before proceeding 
with the following operations it is necessary to proceed 
with the adjustment of the dead band (see attachment 
NI-704). 
Increase the pressure in the circuit up to the desired set 
point value for the first microswitch. 
Use a wide bladed screwdriver, as indicated on the 
adhesive plate, until the relative lamp turns on (or turns 
off). 
- If the instrument is equipped with only one contact the 
calibration is complete. 
- If it is equipped with two contacts continue in the 
following manner. 
Vary the pressure until the desired set point value for 
the second microswitch is reached (fig. 3). 
Act on the adjustment screw of the second contact. 
Repeat former operations on the first contact, then 
operations on the second contact, until the required set 
point precision is obtained. This is necessary due to the 
reciprocal influence which the microswitches have on 
the sensitive element of the instrument. 
Check the calibration values (varying the pressure in the 
circuit accordingly) and record them on the adhesive 
plate using a pen with indelible ink. 
5.3 FINAL OPERATIONS 
Disconnect the instrument from the calibration circuit. 
Weatherproof pressure switches (Series APW) 
Take the cover, ensure that the sealing gasket is 
correctly fitted into its seat, and insert the cover onto the 
case, with the blocking gap positioned in 
correspondence to the blocking bracket. Turn the cover 
clockwise closing it tightly. Mount the adjustment screw 
access plate, then the blocking device as in Fig.4. 
Explosionproof pressure switches (Series APA). 
Insert the closure plugs of the adjustment screw access 
holes, block them using the internal device and if 
necessary seal them with plumbing. Screw on the cover 
and block it using the headless screw with which it is 
equipped (Fig. 5). 
Replace the supplied protection caps on the pressure 
attachment and cable conduit. 
IMPORTANT: the protection caps should only be 
definitively removed during the connection steps (see 
§6). 
6. MOUNTING AND CONNECTIONS
6.1 MOUNTING
Surface mount the instrument by means of the holes 
provided, or pipe mount using the appropriate bracket 
(see Fig. 9). The chosen position must be such that 
vibrations, the possibility of shocks or temperature 
changes are within tolerable limits. The above also 
applies to direct mounting. With gas or vapour process 
fluid, the instrument must be positioned higher than the 
pipe inlet (see Fig. 8). With a liquid process fluid, the 
instrument can be positioned higher or lower, 
indifferently (see Fig. 8 and 9). ). In this case, during set 
point calibration the negative or positive head must be 
taken into account (distance h in fig.8 and 9). 
6.2 PRESSURE CONNECTIONS 
For a correct installation it is necessary to: 
Mount a shut-off valve with drain (root valve) on the 
process tube to allow the instrument to be excluded and 
the connection tubing to be drained. It is recommended 
that said valve has a capstan-blocking device aimed at 
preventing it being activated casually and without 
authorisation. 
Mount a service valve near the instrument to permit 
possible functional verification on site. It is 
recommended that the service valve is closed with a 
plug to prevent the outlet of the process fluid caused by 
the incorrect use of said valve. 
Mount a three-piece joint onto the threaded attachment 
of the instrument to permit the easy mounting or 
removal of the instrument itself. 
Carry out the connection using a flexible tube in such a 
way that variations in the temperature of the tube itself 
do not force the instrument attachment. 
Ensure that all the pressure connections are airtight. It 
is important that there are no leakage in the circuit. 
Close the root valve and the relative drain device. Close 
the service valve using a safety plug. 
6.3 ELECTRICAL CONNECTIONS
It is recommended to carry out the electrical 
connections according to the applicable standards. In 
case of explosionproof instruments (Series APA) see 
also the Standard EN-60079-14. If the electrical 
connection is carried out in a protected tube, it shall be 
made so that condensate is prevented from entering 
instrument enclosure. 
The arrangement shown in fig. 8 and 9 is therefore 
recommended. 
CAUTION:
fittings used for the electrical connection of
the pressure switch Series APA (explosionproof) shall 
be certified to Standards EN 50014 and 50018, and 
shall guarantee instrument degree of protection (IP65). 
Check that there is no power in the lines. Remove the 
cover and carry out the cabling and connections to the 
terminal block (see Fig. 3). Flexible cables with a 
maximum section of 1.2 mm
2
(16AWG) are
recommended using the pre-insulated fork thimbles. Do 
not touch the adjustment screws and do not bend 
the elastic microswitch supports in order to prevent the 
instrument calibration being altered. Ensure that no 
deposits or wire ends remain inside the case. 
As soon as connection steps are completed, mount the 
cover on and make sure it is tight and blocked. See Fig. 
4 and 5. 
Absolute 
pressure 
