Setting the switching point

In order to adjust the switching point, the gauge should be discon-

nected from the voltage supply and the window unscrewed.

The switches are set via the mark pointer on the dial‘s circum-

ference. The set value of the switching point is adjustable up to

80 % of the scale range.

Electromagnetic compatibility

Reed contacts are electrically passive components, which, when

used in accordance with their intended use, do not affect their

surroundings.

6. Storage

The pressure gauge should be kept in its original packing until

installation. The gauge should be protected from external damage

during storage. Storage temperature: -20 °C ... +60 °C.

Pressure gauges should be protected from dust and humidity.

7. Maintenance and servicing / Repairs

The instruments require no maintenance or servicing.

Checks should be carried out on a regular basis to guarantee the

measuring accuracy of the pressure gauge. Any tests or recalibrati-

ons must be carried out by qualified personnel using the appropri-

ate equipment.

Any residual pressure medium contained in the

pressure element may be hazardous or toxic. This

should be taken into account when handling and

storing pressure gauges which have been removed.

8. Disposal

Dispose of instrument components and packaging materials in

accordance with the respective waste treatment and disposal

regulations of the region or country to which the instrument is

supplied.