3. Technical data and intended use

3. Technical data and intended use

Pressure limitation

Model 21X.20, 23X.50/30: Steady:

full scale value

Fluctuating: 0.9 x full scale value

Short time: 1.3 x full scale value

Model 23X.36: Steady:

full scale value

Fluctuating: 0.9 x full scale value

Short time: overload range

Process connection

In according with the general technical regulations for pressure gauges

(i.e. EN 837-2 „Selection and installation recommendations for pressure
gauges“).

When screwing gauges in, the force required for this must not be applied
through the case, but rather through the spanner flats (using a suitable tool)
provided for this purpose on the square shaft of standard connections.

Temperature effect

When the temperature of the measuring system deviates from the reference
temperature (+0 °C): max. ±0.4 %/10 K of full scale value

IP Ingress protection

Housing (EN 60 59/IEC 60 59)
Model 1.0:

IP 54

Models 3X.X0/36: IP 65

This manual is related to the following products: