l

No power supply

à Check cables for breaks; repair if necessary

l

Operating voltage too low or load resistance too high

à Check, adapt if necessary

?

Current signal greater than 22 mA or less than 3.6 mA

l

electronics module or measuring cell defective

à Exchange instrument or return instrument for repair

In Ex applications, the regulations for the wiring of intrinsically safe
circuits must be observed.

?

E

013

l

no measured value available

9)

à Exchange instrument or return instrument for repair

?

E

017

l

Adjustment span too small

à repeat with modified values

?

E

036

l

no operable sensor software

à Carry out a software update or send the instrument for repair

?

E

041

l

Hardware error

à Exchange instrument or return instrument for repair

Depending on the failure reason and measures taken, the steps
described in chapter "Set up" must be carried out again, if necessary.

8

.3 Instrument repair

You can download a return form (24 KB) in the Internet from our
homepage www.wika.com under the item "S

ervice".

If a repair is necessary, please proceed as follows:

l

Print and fill out one form per instrument

l

If necessary, state a contamination

l

Clean the instrument and pack it damage-proof

l

Attach the completed form and probably a safety data sheet to the
instrument

l

Please contact the agency serving you for the address of the
return shipment

9)

Fault message can also appear if the pressure is higher than the nominal
range.