9
Optimag 082011
Device
Operating voltage range
Vac
115+ 10%
Nominal voltage
Vac
115/ 60 Hz
Frequency
Hz
60
Power consumption (+10%) max. at 115 Vac
W
610
Display
digital
Permissible duration of operation
%
100
Permissible ambient temperature
°C
+5 to +40
Permissible relative humidity
%
80
Protection type acc. to DIN EN 60529
IP 42
Protection class
I
Overvoltage category
II
Contamination level
2
Fuse
T10A/250V
(at nominal voltage
115 V)
Operation at a terrestrial altitude
m
max. 2000
Dimensions (B x T x H)
mm
165 x 275 x 85
Weight
kg
2,5
Motor
Speed range
rpm
50 - 1500
Rated input power
W
16
Setting resolution
rpm
10
Speed variation (no load, nominal voltage,
%
+2
at 1500 rpm,ambient temperature 25 °C)
Stirred quantity max. (H
2
O)
ltr
20
Technical Data
Base plate
Dimensions
mm
ø 135
Material
Al - alloy
Heating
Heating power (-5%/+10%) at nominal voltage W
600
Adjustment and display resolution
K
1
Surface temperature
°C
ambient tempera-
ture to 310
Temperature sensor PT 1000 variation
K
< + (0,15
DIN EN 60751 Kl. A
+0,002xITI)
Temperature variation max. at 100 °C
K
+1,5
Heating plate temperature variation
K
+5
Heating plate control hysteresis, no container, K
+5
center of heating plate at 100 °C
Control hysteresis with temperature
K
+1
sensor PT 1000*
with ETS-D5*
°C
+0,5
with ETS-D6*
°C
+0,2
Adjustable safety circuit
Safety temperature limit (adjustable)
°C
50 - 360 (+10)
* The control accuracy values given were determined using the following set-up:
500 ml water in 600 ml glass beaker, magnetic stirring bar 40, 600 rpm, 50 °C
This product is manufactured by IKA
®
for Chemglass.
Subject to technical changes!