Chemglass CG-1994 User Manual
Page 4

4
Optimag 082011
• Only process media that will not react dangerously to the extra
energy produced through processing. This also applies to any
extra energy produced in other ways, e.g. through light irradiation.
• Please observe the operating instructions for any accessories used.
• Ensure that the external temperature sensor (PT 1000, ETS-D ...)
is inserted in the media to a depth of at least 20 mm.
• The PT 1000 external temperature sensor must always be inser-
ted in the media when connected.
• Safe operation is only guaranteed with the accessories descri-
bed in the ”Accessories” chapter.
• Accessories must be securely attached to the device and cannot
come off by themselves. The centre of gravity of the assembly
must lie within the surface on which it is set up.
• Always disconnect the plug before fitting accessories.
• The appliance can only be disconnected from the mains supply
by pulling out the mains plug or the connector plug.
• The socket for the mains cord must be easily accessible.
• The device will automatically restart in mode B following any
interruption to the power supply.
• It may be possible for wear debris from rotating accessory parts
to reach the material being processed.
• When using PTFE-coated magnetic bars, the following has to be
noted: Chemical reactions of PTFE occur in contact with molten
or solute alkali metals and alkaline earth metals, as well as with
fine powders of metals in groups 2 and 3 of the periodic system
at temperatures above 300 °C - 400 °C. Only elementary fluori-
ne, chlorotrifluoride and alkali metals attack it; halogenated
hydrocarbons have a reversible swelling effect.
(Source: Römpps Chemie-Lexikon and ”Ulmann”, Volume 19)
For protection of the equipment
• The appliance may only be opened by experts.
• The voltage stated on the type plate must correspond to the mains
voltage.
• Do not cover the device, even partially e.g. with metallic plates
or film. This results in overheating.
• Ensure that the base plate is kept clean.
• Protect the appliance and accessories from bumps and impacts.
• Observe the minimum distances between the devices, between
the device and the wall and above the assembly (min. 800 mm),
see fig. 4.
>
1
0
0
m
m
>100 mm
>100 mm
>100 mm
Fig. 4
3957121_Optimag_082011:6 31.08.11 09:14 Seite 4