6.1 Working with the PT1000 external
temperature sensor

Figure 6

The PT1000 external temperature sensor is the

manufacturer’s optional accessory. If the sensor is plugged

in, “Probe” will be shown on the digital display to indicate

the sensor is operating. The setting value of external

temperature sensors and actual temperature are displayed.

The safety circuit controls hotplate temperature.

Comparing with the temperature control of the hotplate, the

external temperature sensor can control the medium’s

temperature more precisely. The external temperature

sensor must be put in sample container, if detect anomalies,

heating will automatically shut down. Please do the

following at this time:

Turn off the power

Ensure the external temperature sensor is positioned

sample container correctly

Turn on the power, set the target temperature and

switch on the heating function

If the unit is still not functioning correctly, please

contact manufacturer/supplier

6.2 Residual heat warningHOT

In order to help prevent the risk of burns from the hotplate,

this instrument has a residual heat warning function. When

the heating function is switched off (and the power cord is

still plugged in) and the heating plate temperature is still

above 50°C, “Hot” will flash to warn that it is unsafe to

touch.. When the hotplate temperature drops to below 50°C,

the display will automatically switch off. To completely

turn off the unit, pull out the plug. When the plug is pulled

out, the residual heat warning function will not function.