Instrument does
not react
Battery discharged
or faulty
Charge battery for at least
5 min without operating,
then only operate with
charging cable attached
until battery is recharged.
Replace battery if needed.
Faulty electronic
component
Send in the instrument
for repair.
Instrument does
not react
Faulty electronic
component
Send in the instrument
for repair.
Instrument does
not react
Unpredicted program
error
Confirm error by press-
ing the ENTER key. The
instrument is reinitialized.
Instrument does
not react
No battery inserted
Insert battery
Battery is defective
Replace battery
Faulty electronic
component
Send in the instrument
for repair.
Tip drips/
instrument not sealed
or volume error
—
Improper tip
Only use quality tips
Tip is not properly
seated
Press tip in firmly/use
other ejector-adjustment
clips.
Piston, nose cone or
seal is contaminated or
damaged
Clean the instrument/
replace the seal.
Oil piston.
Display is dark
Electrostatic discharge Remove and insert
the battery.
Faulty electronic
component
Send in the instrument
for repair.
Troubleshooting
If an error occurs, the instrument display shows "Err" and the error number is also shown. The instrument
will now only react to the ENTER key. Pressing the ENTER key will attempt to restart the instrument.
Therefore, a reference run is automatically requested.
Problem
Error message
Possible cause
Corrective action