Disp mode – BrandTech Scientific Transferpette electronic User Manual
Page 54
3x
1x
1x
+
–
1x
Scroll through the modes
using the arrow keys until
‚DISP‘ appears.
‚Mode‘ continues to blink.
Press the arrow keys (+/-)
to set the volume. Holding
the arrow key down accel-
lerates the rate of change.
‚VOL‘ blinks.
Press the ENTER key.
The display now shows the
new volume setting for the
dispensing steps and the
max. number of steps.
‚Steps‘ blinks.
DISP Mode
The Programs · Settings and Use
Press the MENU key three
times to bring ap the mode
selection menu.
‚Mode‘ blinks.
1.
Bring up the menu
2.
Select DISP
mode
Press the ENTER key.
The Display now shows
‚blo‘ for blow-out.
Press the pipetting key once
to move the piston into the
start position.
The arrow in the display
points upwards (aspiration).
3.
Confirm DISP
mode
4.
Prepare for
dispensing
5.
Set dispensing
step volume
6.
Confirm dispens-
ing step volume
Program for discharging an aspirated liquid in pre-defined steps.
The volume aspirated will be a little bit more than actually needed.
Speed adjustment is described on page 43.
What to do
How to do it
Keys to press
Display readout
54