Revpip mode – BrandTech Scientific Transferpette electronic User Manual
Page 50
3x
1x
1x
1x
1x
➞
➞
Scroll through the modes
using the arrow keys until
‚revPIP‘ appears.
‚Mode‘ continues to blink.
Press the pipetting key
once. The volume aspirated
will be a little bit more than
set.
To discharge the measured
amount of liquid, press the
pipetting key once. The
arrow in the display points
downwards (discharge).
Some liquid will remain
in the tip.
revPIP Mode
The Programs · Settings and Use
Press the MENU key three
times to bring up the mode
selection menu.
‚Mode‘ blinks.
1.
Bring up the menu
2.
Select revPIP
mode
Press the ENTER key.
The Display now shows ‚blo‘
for blow-out.
Press the pipetting key once
to move the piston into the
start position.
The arrow in the display
points upwards (aspiration).
3.
Confirm revPIP
mode
4.
Prepare for
pipetting
5.
Aspirate liquid
6.
Discharge liquid
in the revPIP
mode
Program for pipetting of liquids with high viscosity, vapor pressure or that tend to foam.
Volume and speed adjustments are described on pages 42 and 43.
What to do
How to do it
Keys to press
Display readout