Easy calibration – BrandTech Scientific Transferpette electronic User Manual
Page 59
1x
1x
>3s
English
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
‚on‘. ‚CAL‘ blinks.
Press one of the arrow keys
to deactivate the CAL mode.
The display changes from
‚on‘ to ‚off‘. ‚CAL‘ continues
to blink.
Press the ENTER key.
The CAL symbol disap-
pears. The instrument has
now been reverted to factory
default setting.
The continually displayed CAL symbol refers to a previously made
adjustment.
1.
Bring up the CAL
mode
2.
Deactivate CAL
mode
3.
Revert to factory
setting
The instrument is permanently adjusted for watery solutions,
but it can also be set for solutions with varying density, vis-
cosity and temperature. The Transferpette
®
electronic can be
adjusted in every mode, with the exception of the GEL mode.
When the Transferpette
®
electronic is adjusted, a volume
offset is performed, which means that the volume is changed
across the entire volume range of the pipette by the same
amount. It is recommend that the adjustment be performed
at 50% of the nominal volume.
Easy Calibration
What to do
How to do it
Keys to press
Display readout
Revert to factory default settings
Important:
59
Note: