beautypg.com

Operation – Nor-Lake General Purpose Lab Ref-Fzr User Manual

Page 2

background image

2

05/07 138708

Choose a location near a grounded electrical outlet. Do not install the refrigerator-freezer where the

temperature will drop below 55°F (13°C) or rise above 110°F (43°C). The compressor will not be

able to maintain proper temperatures inside the refrigerator-freezer.
Allow adequate space around the unit for good air circulation.

A minimum of 3/8 inch (10 mm)

space on the sides and top and 1 inch (25 mm) space at the back of the refrigerator-freezer is

required for adequate air circulation.

Do not block the toe grille on the lower front of the refrigerator-freezer. Sufficient air circulation is

essential for proper operation of the unit.
Note:

The exterior walls of the refrigerator may become quite warm as the compressor works to

transfer heat from the interior. Temperatures as much as 30°F (-1°C) warmer than room

temperature can be expected. It is particular in hotter climates to allow adequate space for air

circulation around the refrigerator.

ELECTRICAL

Check the proposed outlet to be used to insure that the voltage, phase, and current carrying

capacity of the circuit from the electrical panel correspond to the requirements of the cabinet.
NEVER

use an extension cord or adapter plug to wire any unit. Refer to the serial tag for all

pertinent electrical information.

The refrigerator-freezer must be plugged into its own 115 volt, 60 Hz, single phase outlet. The

power cord of the refrigerator is equipped with a three-prong grounding plug for protection against

shock hazards. It must be plugged into a properly grounded three-prong receptacle. The

receptacles must be installed in accordance with local codes and ordinances.

If voltage varies by more than 10%, refrigerator performance may be affected. Operating the

refrigerator with insufficient power can damage the compressor.

To prevent the refrigerator from being accidentally turned off, do not plug the unit into an outlet

controlled by a wall switch or pull cord.

Do not pinch, knot or bend the cord in any manner.


Observe all Warning Labels. Disconnect power supply to eliminate injury from electrical
shock or moving parts when servicing equipment.

OPERATION


Cool Down Period

For safe storage of product, allow a minimum of 12 hours for the refrigerator to cool down

completely. The refrigerator will run continuously for the first several hours.


Temperature Control

The temperature control is located inside the refrigerator. When first energizing the refrigerator-

freezer, move the refrigerator and freezer controls to ▼ which is the recommended initial setting.

Adjust the controls after 24 hours if required. Adjust the temperatures gradually. Move the control in

small increments and allow the temperature to stabilize between setting changes.
For colder temperatures, adjust the control knob towards Colder.
For warmer temperatures , adjust the control knob towards Cold.

Turning the refrigerator control will change temperatures in both the refrigerator and the freezer. If

the temperature control is turned to a colder setting, the freezer control may require adjustment to a

warmer setting. Turning the freezer temperature control will change only the freezer temperature.

To maintain required temperatures, a fan circulates air in the refrigerator and freezer compartments.

Do not block cold air vents with product.