beautypg.com

Nor-Lake General Purpose Lab Ref-Fzr User Manual

Nor-Lake Refrigerators

background image

LABORATORY REFRIGERATOR-

FREEZER

Installation, Operation and

Maintenance Instructions

INSPECTION

When the equipment is received, all items should be carefully checked against the bill of lading to

insure all crates and cartons have been received. All units should be inspected for concealed

damage by uncrating the units immediately. If any damage is found, it should be reported to the

carrier at once, and a claim should be filed with the carrier. This equipment has been inspected and

tested at the manufacturing facility and has been crated in accordance with transportation rules and

guidelines. Manufacturer is not responsible for freight loss or damage.

Before connecting the refrigerator to the power supply, let it stand for approximately two hours to

reduce the possibility of malfunctions in the cooling system due to transport handling

INSTALLATION

GENERAL

After the unit crate and crate base have been removed, ensure that the cabinet is level. All four

corners of the unit must rest firmly on a solid floor. The cabinet is equipped with adjustable front

rollers to level the unit. Remove the toe grill as shown below and use a flat bladed screwdriver or

3/8” socket wrench to adjust the front rollers (see the illustration below). Raise the front of the

cabinet enough so that the doors will close freely when opened halfway. The cabinet should slope

¼” to ½” from front to back. Then level the cabinet from side to side.
Note:

It is extremely important that the refrigerator be level in order to function properly. If the

refrigerator is not properly leveled during installation, the door may be misaligned and not close or

seal properly causing cooling, frost or moisture problems.

1

05/07 138708