Agilent Technologies System DC N5700 User Manual
Page 121
Service Appendix C
Series N5700 User’s Guide
121
Execution
Errors
(continued)
−231
Data questionable
The measurement accuracy is suspect.
−232
Invalid format
The data format or structure is inappropriate.
−233
Invalid version
The version of the data format is incorrect to the instrument.
−240
Hardware error
The command could not be executed because of a hardware problem with the instrument.
−241
Hardware missing
The command could not be executed because of missing hardware, such as an option.
−260
Expression error
An expression program data element related error occurred.
−261
Math error in expression
An expression program data element could not be executed due to a math error.
Query
Errors
(these errors set Standard Event Status register bit #2)
−400
Query Error
Generic error query
−410
Query INTERRUPTED
A condition causing an interrupted query error occurred.
−420
Query UNTERMINATED
A condition causing an unterminated query error occurred.
−430
Query DEADLOCKED
A condition causing a deadlocked query error occurred.
−440
Query UNTERMINATED after indefinite response
A query was received in the same program message after a query indicating an
indefinite response was executed.