Appendix C Service
120
Series N5700 User’s Guide
Command
Errors
(continued)
−141
Invalid character data
Either the character data element contains an invalid character, or the element is not valid.
−144
Character data too long
The character data element contains more than 12 characters.
−148
Character data not allowed
A discrete parameter was received, but a string or numeric parameter was expected.
−150
String data error
Generic string data error
−151
Invalid string data
An invalid character string was received. Check that the string is enclosed in quotation marks.
−158
String data not allowed
A character string was received, but is not allowed for this command.
−160
Block data error
Generic block data error
−161
Invalid block data
The number of data bytes sent does not match the number of bytes specified in the header.
−168
Block data not allowed
Data was sent in arbitrary block format but is not allowed for this command.
−170
Expression error
Generic expression error
−171
Invalid expression data
The expression data element was invalid.
−178
Expression data not allowed
Expression data element was sent but is not allowed for this command.
Execution
Errors
(these errors set Standard Event Status register bit #4)
−200
Execution error
Generic syntax error
−220
Parameter error
A data element related error occurred.
−221
Settings conflict
A data element could not be executed because of the present instrument state.
−222
Data out of range
A data element could not be executed because the value was outside the valid range.
−223
Too much data
A data element was received that contains more data than the instrument can handle.
−224
Illegal parameter value
An exact value was expected but not received.
−225
Out of memory
The device has insufficient memory to perform the requested operation.
−226
Lists not same length
One or more lists are not the same length.
−230
Data corrupt or stale
Possible invalid data. A new reading was started but not completed.