Test record form – agilent n5742a and n5762a – Agilent Technologies System DC N5700 User Manual
Page 102
Appendix B Verification and Calibration
102
Series N5700 User’s Guide
Test Record Form – Agilent N5742A and N5762A
Agilent N5742A and N5762A
Report No _______________
Date __________________
Description
Model
Minimum Specs.
Results
Maximum Specs.
Voltage Programming & Readback
Minimum Voltage Vout
Measurement Readback
High Voltage Vout
Measurement Readback
Both
Both
Both
Both
0 mV
Vout 8 mV
7.992 V
Vout 16 mV
__________
__________
__________
__________
+20 mV
Vout + 8 mV
8.008 V
Vout + 16 mV
CV Load Effect
Both
2.8 mV
__________
+ 2.8 mV
CV Source Effect
Both
2.8 mV
__________
+ 2.8 mV
CV Ripple and Noise
peak-to-peak
rms
Both
Both
N/A
N/A
__________
__________
60 mV
8 mV
Transient Response
Voltage @ 1.5ms
Both
40 mV
__________
+ 40 mV
Current Programming & Readback
Minimum Current Iout
Measurement Readback
High Current Iout
Measurement Readback
N5742A
N5762A
N5742A
N5762A
N5742A
N5762A
N5742A
N5762A
0 mA
0 mA
Iout 270 mA
Iout 495 mA
89.82 A
164.67 A
Iout 360 mA
Iout 660 mA
__________
__________
__________
__________
__________
__________
__________
__________
+ 452 mA
+ 827 mA
Iout + 270 mA
Iout + 495 mA
90.18 A
165.33 A
Iout + 360 mA
Iout + 660 mA
CC Load Effect
N5742A
N5762A
23 mA
38 mA
__________
__________
+ 23 mA
+ 38 mA
CC Source Effect
N5742A
N5762A
11 mA
18.5 mA
__________
__________
+ 11 mA
+ 18.5 mA
WARNING
Return the voltage and current settings to zero when verification is completed
Test Description
N5742A Settings
N5762A Settings
Voltage Programming & Readback, Min Voltage
≤16 mV, 360 mA
≤16 mV, 660 mA
Voltage Programming & Readback, High Voltage
8 V, 360 mA
8 V, 660 mA
CV Load Effect, Source Effect, Ripple and Noise
8 V, 90 A
8 V, 165 A
Transient Response
8 V, from 9 A to 81A
8 V, from 16.5 A to 148.5A
Current Programming & Readback, Min Current
360 mA, 8 V
660 mA, 8 V
Current Programming & Readback, High Current
90 A, 8 V
165 A, 8 V
CC Load Effect, Source Effect
90 A, 8 V
165 A, 8 V
Load Requirements
N5742A
N5762A
Current shunt
0.01 100 A
0.001 300 A
Agilent N3300 Electronic load modules
2 – N3305A
3 – N3305A
Fixed Resistor for CV Ripple and Noise
0.09 1kW
0.04 2kW