Wiring, Example programs, 1. wiring for example programs – Campbell Scientific SI-111 (IRR-P) Precision Infrared Radiometer User Manual

Page 9

background image

SI-111 Precision Infrared Radiometer

4. Wiring

Target Temperature:
Red

Differential High

Black Differential

Low

Clear Analog

Ground

Sensor Body Temperature:
Green Single-Ended

Blue Analog

Ground

White

Voltage Excitation

5. Example Programs

The example datalogger programs measure the SI-111’s thermistor to obtain
the SI-111 sensor body temperature and measure the SI-111’s thermopile to
obtain the target-to-sensor body temperature difference.

After measuring the thermopile and thermistor outputs, the sensor body
temperature is used to reference the target temperature.

Wiring for the example programs is shown in TABLE 5-1. The actual channels
used need to be adjusted for the actual installation and application.

Coefficients used to calculate the slope (m) and intercept (b) are
specific to individual SI-111 sensors. The unique coefficients for
each individual sensor are provided on the calibration sheet
shipped with the sensor.

NOTE

TABLE 5-1. Wiring for Example Programs

Sensor/Lead Description

CR10X

CR1000

SI-111 Thermopile

Target Temp

Red

Diff.

High

2H

2H

Black

Diff.

Low

2L

2L

Clear

Analog

Ground

AG

SI-111 Thermistor

Sensor Temp

Green

SE

1

1

Blue

Analog

Ground

AG

White

Excitation

E1

VX1 or
EX1

3