SI-111 Precision Infrared Radiometer

;{CR10X}

*Table 1 Program
01: 1

Execution

Interval

(seconds)


;Instruction string to measure the resistance of the thermistor and calculate the sensor body
;temperature. See the Instruction Manual for Campbell Scientific Model 109 Temperature Probe for
;details.

1: AC Half Bridge (P5)
1: 1

Reps

2: 25

2500 mV 60 Hz Rejection Range ;the range should at least match the excitation

3: 1

SE

Channel

4: 1

Excite

all

reps

w/Exchan

1

5: 2500

mV

Excitation

6: 1

Loc

[

mV_thrm

]

7: 1.0

Mult

8: 0.0

Offset


2: Z=1/X (P42)
1: 1

X Loc [ mV_thrm ]

2: 2

Z Loc [ 1_mV_thrm ]


3: Z=X+F (P34)
1: 2

X

Loc

[

1_mV_thrm

]

2: -1.0

F

3: 3

Z Loc [ 2_mV_thrm ]


4: Z=X*F (P37)
1: 3

X

Loc

[

2_mV_thrm

]

2: 24900 F
3: 4

Z Loc [ R_thrm ]


5: Z=LN(X) (P40)
1: 4

X

Loc

[

R_thrm

]

2: 5

Z Loc [ InR_thrm ]


6: Z=X*F (P37)
1: 5

X

Loc

[

InR_thrm

]

2: 0.001

F

3: 6

Z Loc [ Scaled_R ]


7: Polynomial (P55)
1: 1

Reps

2: 6

X Loc [ Scaled_R ]

3: 7

F(X) Loc [ SH_Coeffs ]

4: .001129 C0
5: .234108 C1
6: 0.0

C2

7: 87.7547 C3
8: 0.0

C4

9: 0.0

C5


8: Z=1/X (P42)
1: 7

X Loc [ SH_Coeffs ]

2: 8

Z Loc [ SB_Temp_K ]