;Target temperature calculation based on m and b coefficients.

15: Z=F x 10^n (P30)

1: 0.4

F

2: 1

n,

Exponent

of

10

3: 14

Z Loc [ Exponent1 ]

16: Z=F x 10^n (P30)

1: 0.025

F

2: 1

n,

Exponent

of

10

3: 15

Z Loc [ Exponent2 ]

17: Z=X^Y (P47)

1: 8

X Loc [ SB_Temp_K ]

2: 14

Y Loc [ Exponent1 ]

3: 16

Z Loc [ 1_SB_4Pow ]


18: Z=X*Y (P36)
1: 11

X Loc [ mV_tpile ]

2: 12

Y Loc [ m_slope ]

3: 17

Z Loc [ 2_mVxm ]


19: Z=X+Y (P33)
1: 16

X

Loc

[

1_SB_4Pow

]

2: 17

Y

Loc

[

2_mVxm

]

3: 18

Z Loc [ 3_Sum1 ]


20: Z=X+Y (P33)
1: 13

X

Loc

[

b_inter

]

2: 18

Y

Loc

[

3_Sum1

]

3: 19

Z Loc [ 4_Sum2 ]


21: Z=X^Y (P47)
1: 19

X

Loc

[

4_Sum2

]

2: 15

Y Loc [ Exponent2 ]

3: 20

Z Loc [ T_Temp_K ]


22: Z=X+F (P34)
1: 20

X Loc [ T_Temp_K ]

2: -273.15 F
3: 21

Z Loc [T_Temp_C ]


;Output average values once an hour

23: If time is (P92)
1: 0

Minutes (seconds -- ) into a

2: 60

Interval (same units as above)

3: 10

Set Output Flag High (Flag 0)


24: Real Time (P77)
1: 1220

Year, Day, Hour/Minute (midnight = 2400)


25: Average (P71)
1: 1

Reps

2: 21

Loc

[

T_Temp_C

]

8