Model HFP01SC Self-Calibrating Soil Heat Flux Plate


15: End (P95)

;Compute in-situ calibration.
;
16: If time is (P92)
1: 39

Minutes (Seconds --) into a

2: 180

Interval (same units as above)

3: 30

Then Do


17: Do (P86)
1: 28

Set Flag 8 Low


18: Beginning of Loop (P87)
1: 0

Delay

2: 6

Loop Count


19: Z=X (P31)
1: 7 --

X Loc [ shf_mV_1 ]

2: 31 --

Z Loc [ mV_end_1 ]


20: Z=X*Y (P36)
1: 37 --

X Loc [ V_Rf_1 ]

2: 37 --

Y Loc [ V_Rf_1 ]

3: 13 --

Z Loc [ cal_1 ]


21: Z=X*F (P37)
1: 13 --

X Loc [ cal_1 ]

2: 128.7

F

3: 13 --

Z Loc [ cal_1 ]


22: Z=X+Y (P33)
1: 19 --

X Loc [ mV_0_1 ]

2: 31 --

Y Loc [ mV_end_1 ]

3: 43

Z Loc [ work ]


23: Z=X*F (P37)
1: 43

X Loc [ work ]

2: .5

F

3: 43

Z Loc [ work ]


24: Z=X-Y (P35)
1: 43

X Loc [ work ]

2: 25 --

Y Loc [ mV_180_1 ]

3: 43

Z Loc [ work ]


25: Z=ABS(X) (P43)
1: 43

X Loc [ work ]

2: 43

Z Loc [ work ]


26: Z=X/Y (P38)
1: 13 --

X Loc [ cal_1 ]

2: 43

Y Loc [ work ]

3: 13 --

Z Loc [ cal_1 ]


27: End (P95)

28: End (P95)

29: End (P95)

End Program

-Input Locations-
1 shf_1 1 1 1
2 shf_2 0 0 0
3 shf_3 0 0 0
4 shf_4 0 0 0