Model HFP01SC Self-Calibrating Soil Heat Flux Plate

;{CR10X}
;

*Table 1 Program
01: 1

Execution Interval (seconds)


;Measure HFP01SC on smallest range.
;
1: Volt (SE) (P1)
1: 6

Reps

2: 22

7.5 mV 60 Hz Rejection Range

3: 1

SE Channel

4: 7

Loc [ shf_mV_1 ]

5: 1

Mult

6: 0

Offset


;Measure HFP01SC on larger range.
;
2: Volt (SE) (P1)
1: 6

Reps

2: 23

25 mV 60 Hz Rejection Range

3: 1

SE Channel

4: 44

Loc [ shf_mV_1a ]

5: 1

Mult

6: 0

Offset


;Load in the factory calibration.
;
3: If (X<=>F) (P89)
1: 13

X Loc [ cal_1 ]

2: 1

=

3: 0

F

4: 30

Then Do


;Factory calibration in W/(m^2 mV) = 1000/sensitivity.
;
4: Z=F (P30)
1: 1

F

;<- Enter the unique calibration for plate 1 here.

2: 0

Exponent of 10

3: 13

Z Loc [ cal_1 ]


5: Z=F (P30)
1: 1

F

;<- Enter the unique calibration for plate 2 here.

2: 0

Exponent of 10

3: 14

Z Loc [ cal_2 ]


6: Z=F (P30)
1: 1

F

;<- Enter the unique calibration for plate 3 here.

2: 0

Exponent of 10

3: 15

Z Loc [ cal_3 ]


7: Z=F (P30)
1: 1

F

;<- Enter the unique calibration for plate 4 here.

2: 0

Exponent of 10

3: 16

Z Loc [ cal_4 ]


8: Z=F (P30)
1: 1

F

;<- Enter the unique calibration for plate 5 here.

2: 0

Exponent of 10

3: 17

Z Loc [ cal_5 ]


9: Z=F (P30)
1: 1

F

;<- Enter the unique calibration for plate 6 here.

2: 0

Exponent of 10

3: 18

Z Loc [ cal_6 ]


10: End (P95)