*Table 2 Program
02: 0

Execution Interval (seconds)


*Table 3 Subroutines

;Calibration routine.
;
1: Beginning of Subroutine (P85)
1: 8

Subroutine 8


;Perform in-situ calibration.
;
2: If time is (P92)
1: 1

Minutes (Seconds --) into a

2: 180

Interval (same units as above)

3: 30

Then Do


3: Z=X (P31)
1: 2

X Loc [ shf_mV ]

2: 4

Z Loc [ mV_0 ]


;Begin heating for calibration.
;
4: Do (P86)
1: 48

Set Port 8 High


;Used to filter data during and after calibration.
;
5: Do (P86)
1: 18

Set Flag 8 High


6: End (P95)

;End site calibration three minutes after calibration started.
;
7: If time is (P92)
1: 4

Minutes (Seconds --) into a

2: 180

Interval (same units as above)

3: 30

Then Do


;Measure voltage across current shunt resistor (10 ohm 1% 0.25 W 50
;ppm/deg C) during calibration. This measurement is used to
;compute power.
;
8: Volt (SE) (P1)
1: 1

Reps

2: 25

2500 mV 60 Hz Rejection Range

3: 2

SE Channel

4: 7

Loc [ V_Rf ]

5: .001

Mult

6: 0

Offset


9: Z=X (P31)
1: 2

X Loc [ shf_mV ]

2: 5

Z Loc [ mV_180 ]


;Turn off the soil heat flux plate heater.
;
10: Do (P86)
1: 58

Set Port 8 Low


11: End (P95)

11