7. Installation
Several configurations are possible when using the EnviroSMART
TM
/ EasyAG
®
Probes with Campbell Scientific dataloggers. Among those configuration are the
following:
Option 1 (Preferred)
Device
Function
Probe
Output scaled frequency
Datalogger
Store scaled frequency, calculate and store water content
Option 2
Device
Function
Probe
Output water content
Datalogger
Store water content
Option 1 allows for use of data in Sentek’s Irrimax Software, which allows for
application of new soil calibrations after data are collected. Option 1 is used for
the purposes of this manual.
Refer to the installation manual published for each probe for information on
access tube installation. Appendix A addresses probe normalization.
8. Probe Wiring
A very small flat-point screw driver is used to open the gates of the EasyAG
®
terminals before insertion of wires. Tighten the gate screws after inserting wires.
8.1 SDI-12 Interface Wiring
FIGURE 1. EasyAG
®
SDI-12 Interface
Probe
Pin
CR10X / CR23X
CR200 Series
SDI-12
3
C1
C1 / SDI-12
Gnd 2 G
or
s
+Vin
1
12 V
SW Battery
5