Appendix A. Normalization and
Function Test

Normalization is the setting of the range over which the electronics is effective.
This range is bounded by the two extremes of air and water.

A.1 Normalization

Connect +12 Volts DC and ground to the green connector at the top of the
probe as shown in Figure A-1.

FIGURE A-1. SDI-12 Interface Power Connection

Connect the Intelligent Probe Utility Cable (P/N SEN06020) to the TTL port
near the top of the probe at the location indicated in Figure A-2.

FIGURE A-2. SDI-12 Interface TTL Connection

Connect the other end of the utility cable to the serial port on a PC. This may
require a 9 to 25 pin serial adaptor. CSI serial cable (P/N 7026) is available as
an adaptor.

A-1