beautypg.com

4 cs616 sample edlog program 4, Cs616 sample edlog program 4 – Campbell Scientific CS616 and CS625 Water Content Reflectometers User Manual

Page 25

background image

CS616 and CS625 Water Content Reflectometers

;{CR10X}
;
*Table 1 Program
01: 3600

Execution Interval (seconds)


1: Batt Voltage (P10)
1: 7

Loc [ batt ]


2: CS616 Water Content Reflectometer (P138)
1: 3

Reps

2: 1

SE Channel

3: 6

C6 is first of sequential Control Ports used

4: 1

Loc [ period_1 ]

5: 1.0

Mult

6: 0.0

Offset


3: Polynomial (P55)
1: 3

Reps

2: 1

X Loc [ period_1 ]

3: 4

F(X) Loc [ vwc_1 ]

4: -0.0663 C0
5: -0.0063 C1
6: 0.0007 C2
7: 0.0

C3

8: 0.0

C4

9: 0.0

C5


;
4: If time is (P92)
1: 0000

Minutes (Seconds --) into a

2: 240

Interval (same units as above)

3: 10

Set Output Flag High (Flag 0)


5: Average (P71)
1: 4

Reps

2: 4

Loc [ vwc_1 ]


*Table 2 Program
02: 0.0000 Execution Interval (seconds)

*Table 3 Subroutines

End Program

-Input Locations-
1 period_1 5 1 1
2 period_2 9 1 1
3 period_3 17 1 1
4 vwc_1 5 1 1
5 vwc_2 9 1 1
6 vwc_3 17 1 1
7 batt 1 0 1

6.4.5.4 CS616 Sample Edlog Program 4

Measure soil temperature with 107 probe and CS616 output period every four

hours. Correct CS616 output period for temperature and write result to

datalogger final storage.

19