beautypg.com

2 using edlog instruction 27 (period averaging), Using edlog instruction 27 (period averaging) – Campbell Scientific CS616 and CS625 Water Content Reflectometers User Manual

Page 20

background image

CS616 and CS625 Water Content Reflectometers

SE Chan: Enter the single-ended analog channel where the sensor’s green wire

is connected. When Reps is greater than 1, this entry is the channel for the first

CS616.

Port: Enter an integer to specify the control port that will be used to enable the

CS616 sensor. For the CR510, only C1 can be used to enable a CS616.

Code

Control Port Option

X

X specifies the first control port that will be used. Subsequent

repetitions will be enabled with the next sequential control port.

Control port 1 follows control port 8 in a sequence.

1X

All repetitions will be enabled with a single specified control port.

Simultaneously enabling several CS616s can result in exceeding

power supply capacity.

Loc: Enter the input location that will store the period measurement. The

period output is in microseconds.

An Input Location is a place in the datalogger's memory where a measurement

is temporarily stored until it is used in intermediate storage, output, or

overwritten. An input location is reserved for each measurement in the

datalogger program. Additional locations can be created by the user.

Mult: A factor that the input location value is multiplied by. Enter a 1 for

probe output period in microseconds.

A multiplier is often used for calibration or to convert the input location value

to different units.

Offset: A constant that is added to the input location value. An offset of 0 has

no effect on the input location value.

An offset is often used for calibration or to convert the data to different units.

With a multiplier of 1 and an offset of 0, P138 returns the period in

microseconds.

6.4.4.2 Using Edlog Instruction 27 (Period Averaging)

See Section 6.4.5.2, CS616 Sample Edlog Program 2, for example on using

Instruction 27.

1: Period Average (SE) (P27)

1: 1

Reps

2: 00

Range Option

3: 00

SE Channel

4: 0000

No. of Cycles

5: 0000

Timeout (units = 0.01 seconds)

6: 0000

Loc [ _________ ]

7: 1.0

Mult

8: 0.0

Offset

14