8. Maintenance and Calibration

On a monthly basis the level of the pyranometer should be checked. Any dust

or debris on the sensor head should be removed. The debris can be removed

with a blast of compressed air or with a soft bristle, camel hair brush.

Handle the sensor carefully when cleaning. Be careful not

to scratch the surface of the sensor.

Recalibrate the CS300 every three years. Obtain an RMA number before

returning the CS300 to Campbell Scientific, Inc. for recalibration.

9. Troubleshooting

Symptom: –9999 or radiation values around 0

1. Check that the sensor is wired to the single-ended channel specified by the

measurement instruction.

2. Verify that the range code is correct for the datalogger type.

3. Disconnect the sensor leads from the datalogger and use a DVM to check

the voltage between the red (+) and the black (–) wires. The voltage

should be 0 to 200 mV for 0 to 1000 Wm

–2

radiation. No voltage

indicates a problem with either the photodiode or the shunt resistor, both

of which are potted in the sensor head and cannot be serviced.

Symptom: Incorrect solar radiation

1. Make sure the top surface of the sensor head is clean, and that the sensor

is properly leveled.

2. Verify that the range code, multiplier and offset parameters are correct for

the desired engineering units and datalogger type.

Jumps of 3 to 6 Wm

–2

are typical of CR200(X) measurements, due

to the 0.6 mV CR200(X) resolution and the 0.2 mV/Wm

–2

CS300

sensitivity.