7.4 Compensating Measurements for Barometric Pressure

The CRS45X pressure transducer is unvented; therefore it is monitoring sealed

gage pressure and should be corrected for barometric pressure.

7.4.1 Using an Additional CRS45X

One simple way to complete this correction is to deploy an additional CRS45X

as a reference sensor. For this method, in the Select Field Application window,

choose Barometer Pressure Reference (see Section 7.3, Configure CRS45X for

Measurements).

7.4.2 Using a Barometer

The barometric pressure data from a nearby Campbell Scientific weather

station or other Campbell Scientific monitoring platform can also be used to

compensate for barometric pressure. The

Data tab is used to process and

merge the data file from the test sensor with the data file from the barometric

reference sensor. The procedure for doing this follows:

1. Press the browse button next to the

Collected Data File field to browse to

the data files retrieved from the CRS45X.