CRS451/CRS456 Submersible Water-Level Recording Sensor

8. Once the

Data Tables have been configured, the configuration process is

finished and the

Finish button can be selected.

9. Once the configuration is complete, it must be copied to the CRS451

before it will take effect. Click on

Apply Configuration to CRS45X to

start this process.

Before sending the configuration file to the CRS45X, HydroSci may provide a

warning that your start and stop times may need adjusting. If the Start Time is

in the past, the CRS45X will immediately begin logging data but any offset

you may have applied will not be effective. Also, if the Stop Time is in the

past, no data will be logged.

Save data if you have not already done so. Archived data is purged from the

CRS45X when a new configuration file is copied so you will want to collect

any data before sending the configuration file.

This option allows you to save the configuration file for future reference or for

easy use in other similarly deployed sensors.

The CRS451 is calibrated to pressure units in psi. TABLE B-1 contains the

conversion factors used to convert psi to depth of water.