Appendix B. Edlog Programming Examples

26: Z=X (P31)
1: 7

X Loc [ DewPnt_C ]

2: 15

Z Loc [ LowerTmp ]


;Iterative loop to figure out wet-bulb temperature.

27: Beginning of Loop (P87)
1: 0

Delay

2: 25

Loop

Count


28: Z=X (P31)

1: 12

X Loc [ new_wbT ]

2: 11

Z Loc [ old_wbT ]


29: Z=X-Y (P35)

1: 14

X Loc [ UpperTmp ]

2: 15

Y Loc [ LowerTmp ]

3: 12

Z Loc [ new_wbT ]


30: Z=X*F (P37)

1: 12

X Loc [ new_wbT ]

2: 0.5

F

3: 12

Z Loc [ new_wbT ]


31: Z=X+Y (P33)

1: 12

X Loc [ new_wbT ]

2: 15

Y Loc [ LowerTmp ]

3: 12

Z Loc [ new_wbT ]


32: Wet/Dry-Bulb Temp to VP (P57)

1: 9

Pressure Loc [ SVP_kPa ]

2: 4

Dry-bulb Loc [ AirTempC ]

3: 12

Wet-bulb Loc [ new_wbT ]

4: 18

Loc

[

WB_VP_kPa

]


33: Z=X-Y (P35)

1: 11

X Loc [ old_wbT ]

2: 12

Y Loc [ new_wbT ]

3: 16

Z Loc [ Diff_wbT ]


34: Z=ABS(X) (P43)

1: 16

X Loc [ Diff_wbT ]

2: 16

Z Loc [ Diff_wbT ]


35: Z=X-Y (P35)

1: 18

X

Loc

[

WB_VP_kPa

]

2: 10

Y Loc [ VP_kPa ]

3: 13

Z Loc [ DiffVPkPa ]


36: If (X<=>F) (P89)

1: 13

X

Loc

[

DiffVPkPa

]

2: 3

>=

3: 0

F

4: 30

Then

Do

B-4