beautypg.com

Campbell Scientific AM416 Multiplexer User Manual

Page 14

background image

AM416 RELAY MULTIPLEXER

8

EXAMPLE PROGRAMS - GENERALIZED PROGRAM LOOPS FOR THE CR10(X), 21X AND CR7.

21X SAMPLE PROGRAM
*

1

Table 1
Programs

01:

60

Sec.
Execution
Interval

ACTIVATES MULTIPLEXER
01:

P20

Set Port

01:

1

Set high

02:

1

Port
Number

BEGINS MEASUREMENT
LOOP
02:

P87

Beginning
of Loop

01:

0

Delay

02:

16

Loop Count

CLOCK PULSE AND DELAY
03:

P22

Excitation
with Delay

01:

1

EX Chan

02:

1

Delay w/EX
(units=.01
sec)

03:

1

Delay after
EX (units=
.01 sec)

04: 5000

mV
Excitation

04:

USER SPECIFIED

MEASUREMENT
INSTRUCTION

ENDS MEASUREMENT
LOOP
05:

P95

End

DEACTIVATES MULTIPLEXER
06:

P20

Set Port

01:

0

Set low

02:

1

Port
Number

CR7 SAMPLE PROGRAM
*

1

Table 1
Programs

01:

60

Sec.
Execution
Interval

ACTIVATES MULTIPLEXER
01:

P20

Set Port

01:

1

Set high

02:

1

EX Card

03:

1

Port No.

BEGINS MEASUREMENT
LOOP
02:

P87

Beginning
of Loop

01:

0

Delay

02:

16

Loop Count

CLOCK PULSE AND DELAY
03:

P22

Excitation
with Delay

01:

1

EX Card

02:

2

EX Chan

03:

1

Delay w/EX
(units=.01
sec)

04:

1

Delay after
EX (units=
.01 sec)

05: 5000

mV
Excitation

04:

USER SPECIFIED

MEASUREMENT
INSTRUCTION

ENDS MEASUREMENT
LOOP
05:

P95

End

DEACTIVATES MULTIPLEXER
06:

P20

Set Port

01:

0

Set low

02:

1

EX Card

03:

1

Port No.

CR10(X) SAMPLE PROGRAM
*

1

Table 1
Programs

01:

60

Sec.
Execution
Interval

ACTIVATES MULTIPLEXER
01:

P86

Do

01:

41

Set high
Port 1

BEGINS MEASUREMENT
LOOP
02:

P87

Beginning
of Loop

01:

0

Delay

02:

16

Loop Count

CLOCK PULSE
03:

P86

Do

01:

72

Pulse Port
2

DELAY

P22

Excitation
with Delay

01:

1

EX Chan

02:

0

Delay w/EX

03:

1

Delay after EX

04:

0

mV
Excitation

04:

USER SPECIFIED

MEASUREMENT
INSTRUCTION

ENDS MEASUREMENT
LOOP
05:

P95

End

DEACTIVATES MULTIPLEXER
06:

P86

Do

01:

51

Set low
Port 1

FIGURE 6. Example Program Loops for CR10(X), 21X and CR7 Dataloggers.