4WFB120, 4WFB350, 4WFB1K 4 Wire Full Bridge Terminal Input Modules

5.1 Edlog

Dataloggers that use Edlog include CR510, CR10(X), 21X, and CR7. The
Edlog instruction that measures strain gages is Instruction 6 – Full Bridge.

The Input Locations assignments used in CR10(X), 21X, and CR7 Examples
are listed in Table 5-1.

TABLE 5-1. Input

Locations Used in

CR10(X), 21X, and CR7

Examples

Addr Name
1 mVperV
2 mVperV_0
3 Vr
4 uStrain
5 Count
6 GF
7 _4e6
8 Mult
9 1_2Vr
10 Vr_1_2Vr

5.1.1 CR10(X)

;{CR10X}
;
*Table 1 Program
01: 1

Execution

Interval

(seconds)


;Other measurements could be inserted here or before the Output section

1: If Flag/Port (P91)

;On the first execution (Flag 1 is low)

1: 21

Do if Flag 1 is Low

;or when user sets Flag 1 low

2: 1

Call

Subroutine

1

;call the zeroing subroutine


2: Full Bridge (P6)

;Measure the strain gage

1: 1

Reps

2: 22

± 7.5 mV 60 Hz Rejection Range

3: 1

DIFF

Channel

4: 1

Excite all reps w/Exchan 1

5: 2500

mV

Excitation

6: 1

Loc [ mVperV ]

7: 1

Mult

8: 0

Offset

4