Campbell Scientific 255-100 Novalynx Analog Output Evaporation Gauge User Manual
Page 10

255-100 Novalynx Analog Output Evaporation Gauge
DataTable (Hourly,1,-1)
DataInterval
(0,60,min,10)
Sample
(1,PanLevel,FP2)
Minimum
(1,batt_volt,FP2,0,False)
Sample
(1,PTemp,FP2)
EndTable
BeginProg
Scan
(60,Sec,0,0)
PanelTemp
(PTemp,250)
Battery
(Batt_volt)
'Make five measurements then average the readings to reduce error from ripples in the water
BrHalf
(Leveltemp(),5,mV2500,-1,Vx1,5,2500,True ,20000,_60Hz,PanMultiplier,PanOffset)
AvgSpa
(PanLevel,5,Leveltemp(1))
'Call
data
tables
CallTable
Hourly
NextScan
EndProg
CR200-series Example Program
'CR200 Program to measure 255-100 Evaporation Gauge
Public Batt_volt
Public PanLevel
Const PanMultiplier = 1
'Adjust Pan Multiplier and PanOffset
Const PanOffset = 0
'according to sensor calibration
DataTable (Hourly,1,-1)
DataInterval
(0,60,min)
Sample
(1,PanLevel)
Minimum
(1,Batt_volt,0,0)
EndTable
BeginProg
Scan
(60,Sec)
Battery
(Batt_volt)
ExDelSE
(PanLevel,1,1,Ex1,mV2500,500,PanMultiplier,PanOffset)
CallTable
Hourly
NextScan
EndProg
CR10X Example Program
;This CR10X program averages five measurements that occur 0.1 seconds apart to minimize noise
;caused by waves in the stilling well.
01: If time is (P92)
1: 0
Minutes (Seconds --) into a
2: 60
Interval (same units as above)
3: 30
Then
Do
6