9.2 Maintenance

The 109SS probe requires minimal maintenance. Periodically check cabling

for signs of damage and possible moisture intrusion.

9.3 Calibration

Calibration of the 109SS probe is not necessary unless the application requires

removal of the thermistor interchangeability offset described in Section 6,

Specifications. If performing the one point calibration with an Edlog

datalogger, be aware of this precaution:

The value of the offset must be chosen so that the probe outputs the

temperature calculated by the polynomial, not the actual calibration

temperature. For example, a 109SS probe placed in a calibration chamber at 0

°

C outputs 0.1 °C. An Offset argument of –0.08

is required for Edlog

dataloggers because at 0 °C, the polynomial calculates a temperature of 0.02 °C

(Appendix C, Conversion of Thermistor Resistance or Voltage Ratio to

Temperature).