Model 109SS Temperature Probe
8. If LoggerNet, PC400, RTDAQ, or PC200W is running on your PC, and the
PC to datalogger connection is active, you can click
Finish in Short Cut
and you will be prompted to send the program just created to the
datalogger.
9. If the sensor is connected to the datalogger, as shown in the wiring
diagram in step 6, check the output of the sensor in the datalogger support
software data display to make sure it is making reasonable measurements.
5. Overview
The 109SS is a rugged probe that accurately measures soil or water
temperature in a variety of applications. The sensor consists of a thermistor
encased in a stainless-steel sheath. This design protects the thermistor,
allowing the 109SS to be buried or submerged in harsh, corrosive
environments. It can be submerged in water to 45 m (150 ft) or 63 psi. See
Specifications for a complete list of compatible dataloggers.
6. Specifications
Features:
•
Measures soil or water temperature
•
Compatible with AM16/32-series multiplexers
•
Easy to install or remove
•
Durable
•
Compatible with Campbell Scientific CRBasic dataloggers CR200(X)
series, CR800 series, CR1000, CR3000, and CR5000. Also
compatible with Edlog dataloggers CR10(X), CR500, CR510,
CR23X, 21X, and CR7(X)
Sensor Element:
Measurement Specialties Micro-BetaCHIP
Thermistor Probe (MCD) 10K3MCD1
Survival Range:
–50 to 100 °C (thermistor)
–50 to 70 °C (overmolded joint
and cable)
Measurement
Range:
–40 to 70 °C
Time Constant:
31 s in still air
7.5 s in a wind speed of 3 m/s
0.5 s in rolling water or antifreeze
Maximum Cable Length:
1000 ft
Accuracy
1
Worst case:
±0.6 °C over –40 to 70 °C
±
0.49°C over –20 to 70 °C
Interchangeability
Error:
±
0.60 °C at –40 °C
±
0.38 °C at 0 °C
±
0.10 °C at 25 °C
±
0.30 °C at 50 °C
±
0.45 °C at 70 °C
4