Maximum

Steinhart-Hart

Linearization Error:

0.02 °C at –40 °C

FIGURE 6-1. Worst-case probe and measurement errors

FIGURE 6-2. Steinhart-Hart linearization error

1

The overall probe accuracy is a combination of the thermistor interchangeability specification and the

accuracy of the bridge resistor. The Steinhart-Hart equation used in CRBasic instruction

Therm109()

(CRBasic dataloggers) has negligible error. The major error component is the interchangeability

specification of the thermistor. The bridge resistor has a 0.1% tolerance with a 10 ppm temperature

coefficient.