beautypg.com

2 edlog programming, 1 trigger sampler, Edlog programming – Campbell Scientific 10164-L Water Sampler Control Cable for use with Isco and Sigma Autosamplers User Manual

Page 13: Trigger sampler, Program

background image

10164-L Sampler Control Cable for use with Isco and Sigma Autosamplers

TABLE 4-4. Wiring for Example of Inhibiting a

Sigma Sampler Onboard Program

Color

Description

CR800, CR850, CR1000, or CR3000

Purple

Ground

Brown

Control Port

C1

Clear

Shield

For samplers such as the products manufactured by Sigma, the onboard

program is inhibited by setting the control port high.

For example, if the cable is wired as shown in TABLE 4-4, a datalogger

program that includes the following CRBasic instruction will inhibit a Sigma

sampler’s onboard program:

PortSet(1,1)

Above is only a portion of the CRBasic program.

After a Sigma sampler’s program has been inhibited, it can be allowed to run

by setting the port low.

4.2 Edlog Programming

4.2.1 Trigger Sampler

Wire Color

CR10(X)

Purple

G

Yellow

C2

Red

12V

Clear

G

To trigger the sampler, pulse port 2 using a set of instructions such as follows:

Do (P86)
1: 42

Set Port 2 High


; Note: The 50 in the third parameter keeps the
; port high for 0.5 seconds. Some users have
; reported using a delay of 1 sec (100 in
; parameter 3) to ensure reliable triggering of
; the sampler.

Excitation with Delay (P22)
1: 1

Ex Channel

2: 0

Delay W/Ex (units = 0.01 sec)

3: 50

Delay After Ex (units = 0.01 sec)

4: 0

mV Excitation


Do (P86)
1: 52

Set Port 2 Low

NOTE

7