Maximum/Minimum Pressure Capture
The GC31 unit keeps the maximum and minimum pressure level applied to the pressure port
as peak and bottom values respectively, in the internal memory. The peak and bottom values
are displayed while holding the
ᮡ or ᮢ keys respectively. Message pe+ is displayed for one
second and selected Max/Min value is displayed by this operation. Maximum and minimum
values are reset when you reset power to the unit or by the following procedure:
Resetting Maximum value: While holding the
ᮡ key, press the ᮢ key.
Resetting Minimum value: While holding the
ᮢ key, press the ᮡ key.
13. ERROR DISPLAY
An error message and pressure value are alternately displayed when one of the following
errors occurs in the measurement mode (in case of “Out of pressure display range” error, only
error message is displayed). Check the content of error message and taken the action below
immediately.
Error
display
Contents
Actions
Out of pressure display range (Upper limit)
fff
Pressure above 110% FS of pressure had been
applied, or the indicated value exceeded 1999.
Adjust the applied pressure
Out of pressure display range (Lower limit)
within the rated pressure.
-fff
Pressure below –10% FS had been applied, or
the indicated value fell below –1999.
Pressure applied during the zero point
Open the unit to atmosphere
e-0
adjustment,was outside the range by
and perform zero point
± 10% FS.
adjusment.
e-1
Comparator 1 overloaded
Drop load current below
e-2
Comparator 2 overloaded
80mA before use.
Turn power on again to
e12
Comparator 1 & 2 overloaded
recover from error status.