beautypg.com

Anritsu MG3641A User Manual

Page 136

background image

SECTION 6 GPIB

6-30

:FREQuency:SWEep:SPAN

Function

Sets the span frequency for a sweep.

Parameter

= 0.02 Hz to 1040 MHz (MG3641A)

0.02 Hz to 2080 MHz (MG3642A)

Unit

Restriction

None

:FREQuency:SWEep:SPAN?

Function

Reads out the span frequency for a sweep.

Response

Span frequency for a sweep

Restriction

None

:FREQuency:SWEep:STEP:SIZE

Function

Sets the frequency step size for a liner sweep.

Parameter

= 0.01 Hz to 1040 MHz (MG3641A)

0.01 Hz to 2080 MHz (MG3642A)

Unit

Restriction

None

:FREQuency:SWEep:STEP:SIZE?

Function

Reads out the frequency step size for a liner sweep.

Response

Frequency step size for a liner sweep

Restriction

None

:FREQuency:SWEep:STEP:NUMBer

Function

Sets the number of frequency points for a liner sweep.

Parameter

= 2 to 104000000001 (MG3641A)

2 to 208000000001 (MG3642A)

Unit

Restriction

None

:FREQuency:SWEep:STEP:NUMBer?

Function

Reads out the number of frequency points for a liner sweep.

Response

Number of frequency points for a liner sweep

Restriction

None

:FREQuency:SWEep:MODE

Function

Selects the frequency sweep mode.

Parameter

= AUTO

SINGLE
MANUAL

Unit

Restriction

None

:FREQuency:SWEep:MODE?

Function

Reads out the frequency sweep mode.

Response

Frequency sweep mode = AUTO

SINGLE
MANUAL

Restriction

None

This manual is related to the following products: