Anritsu MG3641A User Manual
Page 132
SECTION 6 GPIB
6-26
(3) Device dependent error group
Error codes –300 to –399 indicate that a device error other than command, execution, and query errors occurred in
the device. Bit 3 in the standard event status register of the device is set at occurrence of an error.
Error code
Message
Error detection conditions
–310
System error
A system error occurred.
–311
Save/recall memory lost
The save/recall memory was lost.
–350
Self-test failed
The self-diagnosis failed.
(4) Query error group
Error codes –400 to –499 indicate that a message exchange protocol error occurred in the output buffer control
unit of the device. Bit 2 in the standard event status register of the device is set at occurrence of an error.
Error code
Message
Error detection conditions
–410
Query INTERRUPTED
An interrupt was generated by a new command
before the device sent a response message.
–420
Query UNTERMINATED
No query corresponding to a response message to
be read is sent. The query does not terminate com-
pletely.
–430
Query DEADLOCKED
An attempt was made to buffer a large amount of data
exceeding a storage space.